Specifications |
Insulation Tests |
Test Voltage |
500 Volts DC |
Measurement Range |
0.1 to 1000 MW |
Short Circuit Current |
£6mA |
Accuracy |
±5 percent of Reading |
Automatic Discharge |
1 s/µF |
Continuity Tests |
Measurement Range |
0 to 10 W-, 0 to 10 W+ |
Short Circuit Current |
³200 mA |
Open Circuit Voltage |
4.5 to 6.5 Volts |
Accuracy |
±3 percent full scale |
Voltage Tests |
Measurement Range |
0 - 600 Volts AC |
Frequency |
45 to 400 Hertz |
Accuracy |
3 percent of full scale |
Input Impedance |
300 kW |
Display |
Analog Display |
Rugged taut band suspension |
General Specifications |
Power Source |
Four 1.5 Volts AA Alkaline batteries |
Dielectric Test |
6000 Volts 50/60 Hertz, 1 minute |
Dimensions |
6.6 x 4.2 x 2.2 inches - 167 x 106 x 55 millimeters |
Weight |
1.45 pounds - 650 grams |
Operating Range |
14 to 131 degrees F (-10 to 55 degrees C), 20 - 80 percent RH |
Storage Range |
-40 to 158 degrees F (-40 to 70 degrees C), 10 - 90 percent RH |
Terminals |
"-" Line (black); "+" earth (red); accepts standard 4 millimeter banana plugs |
Safety Reference Specifications |
Safety Ratings |
EN 61010-1, VDE0413-1, VDE0413-4, VDE0100 |
Agency Approvals |
CE Mark, UL |